User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Eagle Flex Timers DATE: 11-Aug-95 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 70 STANDARD: STAND STD-11-2 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # Equipment #Various Type: Eagle Flex Timers # Notes: This is a data only procedure. Use paper procedure and return to the # lab to run this procedure. # # Review Approval:_____________________________ Release Date:____/___/____ 1.001 ASK+ M X 1.002 ASK- R N P F W 1.003 RSLT =Accuracy:ñ30 seconds Repeatability ñ18 seconds. 1.004 RSLT = 1.005 RSLT =Reference Documents:None 1.006 RSLT = 1.007 STD STAND STD-11-2 1.008 DISP Install the UUT into the XP500 Universal Excalibar 1.008 DISP RMS Jammer Special External Test fixture. 1.009 DISP POWER up the UUT and allow it to warm up. 1.010 HEAD {Repeatability Test} 1.011 DISP This is RUN 1 1.011 DISP 1.011 DISP Set the UUT to time out 300 sec (5 MIN) and START. 1.012 MEMI Enter the STANDARD time @ UUT set for 300 SEC 1.013 MEME >1 1.014 DISP This is RUN 2 1.014 DISP 1.014 DISP DO NOT change the UUT setting and repeat. # 1.015 ACC 0.00sec 0.1U 1.015 MEMI Enter the STANDARD time @ UUT set for 300 SEC 1.016 MEME >2 1.017 MATH mem=m[1]-m[2] 1.018 MEMC 0.00sec 18U #! Test Tol 18, Sys Tol 0.1, TUR 180.0000 (>= 4.00). 2.001 HEAD {Range 0-60 min Range} 2.002 DISP Set the UUT to time out 20 minutes and START. # 2.003 ACC 1200sec 1U 2.003 MEMI Enter the STANDARD time @ UUT=1200 seconds. 2.004 MEMC 1200sec 30U #! Test Tol 30, Sys Tol 1, TUR 30.0000 (>= 4.00). 3.001 HEAD 3.002 DISP 3.002 DISP ÿ P R O C E D U R E 3.002 DISP ÿ ÿ COMPLETED! 3.002 DISP #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.